TelePole WR/ Operating & Maintenance Manual Page 11
3.4. Threshold Selection
A threshold value can be selected from a series of 11 threshold values. The value is kept in
memory even if the meter is turned off or if power is lost.
Threshold values:
Exposure rate : 2.00, 5.00, 50.0, 100, 500 [mR/h], 1.00, 10.0, 200, 400, 800, 999 [R/h]
To choose the required threshold value, proceed with the following steps:
3.4.1. Enter threshold mode by pressing the MODE & RESET push buttons simultaneously. The
threshold readings, the SET and RATE segments and the SPK icon will be displayed.
3.4.2. Each short press on the RESET push-button advances the display to the next threshold
value, according to the order described in section 3.4.
3.4.3. To exit the set threshold mode, save and set the new threshold value, press the MODE &
RESET push buttons simultaneously. An audible beep will verify that the new threshold
value was saved.
.