15-00044
Revision 5
April 2008
The publication, translation or reproduction, either party or wholly, of this document are not allowed without our written consent. 9
Optional Accessories
Remote Display Unit (external visual & audible alarm)
Magnet attachments
2.2 DRM-1D/2D Cesium Iodine (TL) Specifications
Detector:
C
sI (Tl) (30 x 30 mm) crystal optically coupled to PMT with 0.5 mm aluminum housing and
solid mu metal shield. With optional automatic background subtraction using the RMV
software.
Measuring Range
1uR/h to 9999 uR/h (0.001 mR/h to 9.999 mR/h)
Overflow Alarm for reading > 9999 uR/h
Overload for field > 1 R/h
Sensitivity: 6.8 cps/uR/h for 137-Cs
Optional : background subtraction via the RMV software
Display range
0.01 mR/h to 10,000 mR/h
Dimensions
DRM-1D: 7" L x 3.5" D x 4.25" W
DRM-2D: 10.25” L x 4.28” D x 4.88 W
Display – LCD Readout:
4
digits for accurate and easy display
Detector failure
Low battery
Overflow
4 Threshold Settings
Audio:
Two internally mounted piezo-electric elements (used for alarm functions)
Accuracy (
137
Cs)
±10 % of reading, within the measuring range
Energy Response
45 KeV to 2 MeV
CsI Sensitivity (
137
Cs):
6.8 cps/µR/h